Fibrinogen
eD-H Order Code
LAB314
Specimen Collection Information
Specimen | Minimum Test Volume | Container | Special Handling |
---|---|---|---|
Blood |
PEDI, PICU, or ICN draws refer to specimen collection job aid. All other Coag tubes must be filled to the line on tube. |
Specimen Transport Temperature
Double spin, separate, and freeze
Reference Ranges / Critical Values
Assay | Gender | Age Range | Normal Range | Critical Values | Units |
---|---|---|---|---|---|
Fibrinogen | All | 0 Minutes – 15 Days | <100 | mg/dL | |
15 Days – 1 Months | 136 -300 | <100 | mg/dL | ||
1 Months – 6 Months | 141 -437 | <100 | mg/dL | ||
6 Months – 1 Years | 148 -367 | <100 | mg/dL | ||
1 Years – 6 Years | 164 -497 | <100 | mg/dL | ||
6 Years – 11 Years | 171 -537 | <100 | mg/dL | ||
11 Years – 18 Years | 168 -529 | <100 | mg/dL | ||
18 Years – 150 Years | 200 -393 | <100 | mg/dL |
Specimen Stability
4 hours
Additional Specimen Instructions
1.8 mL Sodium Citrate tube filled to the line
OR
2.7 mL Sodium Citrate tube filled to the line
Blue top tubes must be filled within 10% of stated volume. A discard tube must be used if a citrate tube is to be drawn using a winged blood collection set. It is important to remove air from the blood collection set to ensure the proper blood volume is obtained in the coag tube.
If needed, 1mL sodium citrate tubes can be prepared by the laboratory. These are a non-vacuum orange screw cap tube with lab prepared anticoagulant. To obtain some of these custom tubes please call the coagulation laboratory at 5-9463.
If the patient has a known hematocrit >55%, please contact the coagulation lab for an adjusted collection tube before drawing the blood.
Interpretive Data
Assay Display | Interpretive Data |
---|---|
Fibrinogen | A fibrinogen level >100 mg/dL is adequate for hemostasis in most patients without underlying bleeding disorders. |
CPT(s)
85384
Performing Lab Section
Hematology