Sign in →

Phospholipase A2 Receptor, Immunofluorescence, Serum

Additional Codes

LAB2596


Ordering Guidance


This test can be used in a complementary fashion with a quantitative enzyme-linked immunosorbent assay to confirm a positive result, especially if it is a low-level titer. Since the immunofluorescence is more sensitive, it can also be used to detect or follow patients with low level antibodies not detected by ELISA.  



Specimen Required


Supplies: Sarstedt Aliquot Tube 5 mL (T914)

Collection Container/Tube:

Preferred: Serum gel

Acceptable: Red top

Submission Container/Tube: Plastic vial

Specimen Volume: 1 mL 

Collection Instructions: Centrifuge and aliquot serum into a plastic vial within 2 hours of collection.


Specimen Type

Serum

Specimen Minimum Volume

0.5 mL

Specimen Stability Information

Specimen Type Temperature Time
Serum Refrigerated (preferred) 14 days
  Frozen  14 days
  Ambient  8 hours

Reference Values

Negative

Day(s) Performed

Monday, Wednesday, Friday

Report Available

3 to 7 days

Specimen Retention Time

7 days

Performing Laboratory

Mayo Clinic Laboratories in Rochester

CPT Code Information

86255

LOINC Code Information

Test ID Test Order Name Order LOINC Value
PLA2I PLA2R, Immunofluorescence, S 82991-1

 

Result ID Test Result Name Result LOINC Value
PLA2I PLA2R, Immunofluorescence, S 82991-1

Forms

If not ordering electronically, complete, print, and send 1 of the following forms with the specimen:

-Kidney Transplant Test Request

-Renal Diagnostics Test Request (T830)